======================================================================= RCS: $Header: a_02_05.txt,v 1.2 94/03/15 15:39:56 hmgr Exp $ ======================================================================= General Information ------------------- This document describes detailed information about the software changes in the A.02.05 version of the HPLC^2D and HPLC^3D ChemStation (DOS Series) software. The revision number convention of the softwares are 'P.AA.II' where : P - Represents the Product number / major revision number link. Versions starting with 'A' are from product numbers that have an identical revision letter. (eg G1304A will always have a revision number A.??.??). It is current APG policy not to change this letter unless the software changes in purpose or function. AA - Represents the major revision number. Changes in this code will typically include significant software enhancements that may require complete revalidation of the system for customers operating in regulated environments. These revisions may also include defect fixes and documentation changes. II - Represents the minor revision number. These revisions correct software defects and are generally isolated in nature probably not requiring revalidation of the system for customers operating in regulated environments. These revisions only include defect fixes. They do not usually have documentation changes. This Revision ------------- This revision is a minor software revision. The changes described are relative to the A.02.04 version of these ChemStations. Further details can be found in the revision history and Software Status Bulletin (SSB) on the supplemental disc of the new version. Operating System ---------------- This revision does not require a change in the operating system. The software runs on MS-DOS Rev. 5.xx and 6.xx and Microsoft Windows version 3.1. Users who design their own report layouts require Microsoft Excel. Revision A.02.05 of the ChemStation is compatible with Microsoft Excel 4.0 and 5.0. Ordering Upgrades ----------------- Upgrades are orderable from HP Analytical Direct (HPAD) in the Americas or through the local representative elsewhere by ordering the following products... G1640A - HPLC 2D single instrument software upgrade for customers upgrading from an existing revision of G1304A to the latest revision. Requires but does not include MS-DOS and MS-Windows. G1641A - HPLC 2D multi instrument software upgrade for customers upgrading from an existing revision of G1305A to the latest revision. Requires but does not include MS-DOS and MS-Windows. G1642A - HPLC 3D single instrument software upgrade for customers upgrading from an existing revision of G1307A to the latest revision. Requires but does not include MS-DOS and MS-Windows. G1643A - HPLC 3D multi instrument software upgrade for customers upgrading from an existing revision of G1309A to the latest revision. Requires but does not include MS-DOS and MS-Windows. G1644A - HPLC 3D data evaluation software upgrade for customers upgrading from an existing revision of G1308A to the latest revision. Requires but does not include MS-DOS and MS-Windows. All these revision upgrades are now US$500 for both US list and reference price. Defect Fixes ------------ The following descriptions highlight the defects that have been fixed in the new revision. For details see the software status bulletin (SSB). ________________________________________________________________________ 1.0 HP 1050 DAD Test Menu SSB Reference - KPR #: B400101451 Description - When selecting the "Instrument" -> "More DAD..." -> "Test" menu, an error message would occur: "undefined symbol: MenuHpceDADTest" This has been fixed in A.02.05. ________________________________________________________________________